The standard calibration scan parameters are:
The standard calibration scan parameters are:
''2theta start: -17 deg
*2theta start: -17 deg


2theta end:  50 deg
*2theta end:  50 deg


step size: 0.0005 deg
*step size: 0.0005 deg


count time: 0.01 sec/step''
*count time: 0.01 sec/step''





Revision as of 16:55, 23 June 2010

A calibration file is needed to merge data collected from each of 11-BM 12 detectors into a single 2theta vs intensity dataset.

The calibration file corrects each of the 12 detectors for 2theta zero shift, wavelength, and scaling factor.

Calibration files are obtained from a 12-histogram GSAS refinement of data collected on a powder sample containing a mixture of NIST SRM Al2O3 & Si.

The standard calibration scan parameters are:

  • 2theta start: -17 deg
  • 2theta end: 50 deg
  • step size: 0.0005 deg
  • count time: 0.01 sec/step


The robot system (aka Snookums) can access 56 samples mounted on individual posts plus 6 magazines of 16 samples each. A diagram for viewing showing the numbering is found here.

Robot Operator Functions