Mail-In Powder Diffraction Measurements

From Ug11bm
Revision as of 15:53, 30 May 2012 by WikiSysop (talk | contribs) (Created page with " == Image Gallery== 305x240px|Sample view through the "upstream" camera. 305x240px|sample view through the video microscope....")
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigationJump to search




Image Gallery

Sample view through the "upstream" camera. sample view through the video microscope.


Two views of a sample in position for data-taking, as seen through the "upstream" camera (left) and the video microscope (right). The reticle lines show the nominal extent of the beam. The actual beam edges extend somewhat farther, so these areas should be kept clear of foreign material (glue, clay, wax, etc.). The sample material must be held in place in the beam -- if it is loose in the capillary, it will move when we spin the capillary (we spin at about 5400 r.p.m.). In addition, there must be no lumps of material at the end of the capillary, since this could cause the capillary to bend or break during spinning, and such a lump may also cause a robot malfunction when it attempts to load or unload the sample.