TEXIO PD18-30AD DC Power Supply

From Ug11bm
Revision as of 18:35, 12 March 2014 by WikiSysop (talk | contribs) (→‎Manuals)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigationJump to search

Notes below are for the TEXIO PD18-30AD DC Power Supply. It is primarily used at 11-BM to provide power for resistive heating coils in the flow cell sample environment.

Manuals

TEXIO_PD18-30AD_DCPowerSupply_Manual

Specs

  • 0 - 30 Amps
  • 0 - 18 V
  • 0 - 450 Watt

Operation & Settings

Front Panel

  • Turn On PD18 Power
  • Adjust Voltage/Current Settings/Limits if needed (see below)
  • For 11-BM resistive heating coils, connect matching Blue-White & Red-Red lead wires (as shown below)
  • For 11-BM resistive heating coils, isolate ground connection (as shown below)
  • ENABLE OUTPUT press OUTPUT button, enabled = red light ON
  • CC (red) and/or CV (green) lights indicate output is limited by Current Limit state (CC) or Voltage Limit state (CV)

Image of Front panel of TEXIO PD18-10AD DC Power Supply

Front panel of TEXIO PD18-10AD DC Power Supply


Back Panel

Image of Back panel of TEXIO PD18-10AD DC Power Supply

  • #1 - 12 are input terminals for remote control and series/parallel operation.
  • For control by external voltage (i.e. SRS temperature control), connect voltage across pins 4 - 5 (as shown below)
  • External voltage should be 0-10 V, with impedance ~ 10 k Ohms

Back panel of TEXIO PD18-10AD DC Power Supply


Voltage / Current Settings

  • Manually set by turning big Current & Voltage Knobs
  • When controlling remotely, these values will set the V & I limits
  • To check setting, press V/I CHECK, then check value and adjust if required.