11-BM has a dedicated Cyberstar Hot Gas Blower. This can be used to heat small diameter capillary samples from ambient temperature to ~ 950 °C
About
Temperature Range and Ramp Rate.
- ambient temperature to ~ 950 °C
- Max ramp rate on heating & cooling is 5 deg C/min (300 C/hour)
Users should plan experiments to incorporate these heating and cooling ramp times.
11-BM Hot Gas Blower Image
Temperature Profile & Alignment
Temperature Calibration
Al2O3 Thermal Expansion
In-situ calibration of the hot gas blower was performed (summer 2012) using the known thermal response of Al2O3 a & c unit cell parameters as a function of temperature.
11-BM calibration results kindly shared by Jae-Hyuk Her, GE Global Research.
J. Appl. Cryst. (1993). 26, 77-81 [ doi:10.1107/S002188989200877X ]
"High-temperature X-ray diffraction: solutions to uncertainties in temperature and sample position"
N. E. Brown, S. M. Swapp, C. L. Bennett and A. Navrotsky