Calibration files are obtained from a 12-histogram GSAS refinement of data collected on a powder sample containing a mixture of NIST SRM Al2O3 & Si. | Calibration files are obtained from a 12-histogram GSAS refinement of data collected on a powder sample containing a mixture of NIST SRM Al2O3 & Si. | ||
The standard calibration scan parameters are: | The standard calibration scan parameters at 30 keV currently are as follows: | ||
*2theta start: -17 deg | *2theta start: -17 deg | ||
Line 12: | Line 12: | ||
*step size: 0.0005 deg | *step size: 0.0005 deg | ||
*count time: 0.01 sec/step | *count time: 0.01 sec/step | ||
= Calibration = | |||
Revision as of 16:56, 23 June 2010
A calibration file is needed to merge data collected from each of 11-BM 12 detectors into a single 2theta vs intensity dataset.
The calibration file corrects each of the 12 detectors for 2theta zero shift, wavelength, and scaling factor.
Calibration files are obtained from a 12-histogram GSAS refinement of data collected on a powder sample containing a mixture of NIST SRM Al2O3 & Si.
The standard calibration scan parameters at 30 keV currently are as follows:
- 2theta start: -17 deg
- 2theta end: 50 deg
- step size: 0.0005 deg
- count time: 0.01 sec/step