=== Al2O3 Thermal Expansion ===   
=== Al2O3 Thermal Expansion ===   


In-situ calibration of the hot gas blower was performed (summer 2012) using the known thermal expansion of Al2O3 (alumina) ''a'' & ''c'' unit cell parameters as a function of temperature.
In-situ calibration of the hot gas blower was performed (summer 2012) using the known ('''Reference 1''') thermal expansion of Al2O3 (alumina) ''a'' & ''c'' unit cell parameters as a function of temperature.


Full diffraction patterns were collected for select (nominal) temperatures between 100 - 950 C at 11-BM using a X.X mm diameter fused quartz capillary load with ??? Al2O3 powder.
Full diffraction patterns were collected for select (nominal) temperatures between 100 - 950 C at 11-BM using a X.X mm diameter fused quartz capillary load with ??? Al2O3 powder.
Line 34: Line 34:




11-BM calibration results kindly shared by Jae-Hyuk Her, GE Global Research.
Note: these 11-BM calibration results a kindly shared by Jae-Hyuk Her, GE Global Research.




J. Appl. Cryst. (1993). 26, 77-81   [ doi:10.1107/S002188989200877X ]
'''Reference 1''' = J. Appl. Cryst. (1993). 26, 77-81
"High-temperature X-ray diffraction: solutions to uncertainties in temperature and sample position"
N. E. Brown, S. M. Swapp, C. L. Bennett and A. Navrotsky
N. E. Brown, S. M. Swapp, C. L. Bennett and A. Navrotsky
dfdfdf


=== Phase Transitions ===   
=== Phase Transitions ===   

Revision as of 21:51, 24 September 2012

About

11-BM has a dedicated Cyberstar Hot Gas Blower. This can be used to heat small diameter capillary samples from ambient temperature to ~ 950 °C

About

Temperature Range and Ramp Rate.

  • ambient temperature to ~ 950 °C
  • Max ramp rate on heating & cooling is 5 deg C/min (300 C/hour)

Users should plan experiments to incorporate these heating and cooling ramp times.


11-BM Hot Gas Blower Image

Hot Air Blower Zoomed Image


Temperature Profile & Alignment

GasBlowerProfile

Temperature Calibration

Al2O3 Thermal Expansion

In-situ calibration of the hot gas blower was performed (summer 2012) using the known (Reference 1) thermal expansion of Al2O3 (alumina) a & c unit cell parameters as a function of temperature.

Full diffraction patterns were collected for select (nominal) temperatures between 100 - 950 C at 11-BM using a X.X mm diameter fused quartz capillary load with ??? Al2O3 powder.

Refined



Note: these 11-BM calibration results a kindly shared by Jae-Hyuk Her, GE Global Research.


Reference 1 = J. Appl. Cryst. (1993). 26, 77-81 N. E. Brown, S. M. Swapp, C. L. Bennett and A. Navrotsky

Phase Transitions

HotGasBlowerTempCalib.png

SiO2

HotGasBlowerSiO2Heating HotGasBlowerSiO2Cooling

PbTiO3

HotGasBlowerPbTiO3