The calibration file corrects each of the 12 detectors for 2theta zero shift, wavelength, and scaling factor.   
The calibration file corrects each of the 12 detectors for 2theta zero shift, wavelength, and scaling factor.   


Calibration files are obtained from a 12-histogram GSAS refinement of data collected on a powder sample containing a mixture of NIST SRM Al2O3 & Si.
Calibration files are obtained from a 12-histogram GSAS refinement of data collected on a powder sample containing a mixture of NIST SRM 676a Al2O3 (~75 wt%) & SRM 640d Si (~25 wt%).


=== Calibration Scan Parameters ===
=== Calibration Scan Parameters ===
 
'''At 30 keV''' - the standard calibration scan parameters currently are as follows:
The standard calibration scan parameters at 30 keV currently are as follows:
*2theta start: -17 deg
*2theta start: -17 deg



Revision as of 17:24, 23 June 2010

A calibration file is needed to merge data collected from each of 11-BM 12 detectors into a single 2theta vs intensity dataset.

The calibration file corrects each of the 12 detectors for 2theta zero shift, wavelength, and scaling factor.

Calibration files are obtained from a 12-histogram GSAS refinement of data collected on a powder sample containing a mixture of NIST SRM 676a Al2O3 (~75 wt%) & SRM 640d Si (~25 wt%).

Calibration Scan Parameters

At 30 keV - the standard calibration scan parameters currently are as follows:

  • 2theta start: -17 deg
  • 2theta end: 50 deg
  • step size: 0.0005 deg
  • count time: 0.01 sec/step

Generating Calibration Files

Users