Difference between revisions of "New Huber slits are not working"

From 26-ID
Jump to navigationJump to search
 
m (1 revision)
 
(No difference)

Latest revision as of 18:30, 13 January 2014

Back to X-Ray Microscopy

New slits have to be calibrated first.


Solution:
1. Manually close slits to zero position
2. In spec, run the command “mot_par(motor,”calibrate”). One can also use “help xiahsc” to find useful information in spec
3. Type the unique serial numbers of the slits in the field 'generic parameters' of the spec configuration file